Rabu, 03 Februari 2010

Pengambilan Gambar CT-scan Advent HD200

Langkah memulai scanning:

1. Letakkan diatas core holder pada gantry sample yang akan di scanning dan posisi diatur dengan bantuan sinar laser.

2. Kemudian kita tekan keypad NEW PATIENT pada operator console, sebabagi berikut:

ENTER PATIENT NAME: TA

ENTER IDENTIFICATION: 01

ENTER PATIENT DIRECTION –HEAD FIRST OR FEET FIRST [ H/F ] : H <>

PATIENT IS IN THE HEAD FIRST POSITION

…..AND IS ORIENTED WITH POSTERIOR SIDE DOWN

IS THE PATIENT DIRECTION AND ORIENTATION CORRECT?

[ Y/N ] : Y

ENTER PATIENT TABLE DIRECTION [+ = OUT, - = IN]: -

PROCEDURE :

TYPE OF ZOOM RECONSTRUCTION

0 = NO ZOOM

1 = CENTERED ZOOM

2 = ZOOM ANYWHERE

ENTER CHOICE [ 0,1,2 ] :

AVAILABLE CONVOLVER FILTERS:

1 = NORMAL

2 = LARGE

3 = LARGE – SHARP

4 = LARGE – SMOOTH

5 = 25 CM

6 = 25 CM - SHARP

7 = 25 CM – SMOOTH

8 = 12 CM

9 = 12 CM - SHARP

10 = 12 CM - SMOOTH

ENTER NORMAL IMAGE FILTER NUMBER: 1

PRESERVE RAW DATA ON DISK?

Y = YES

N = NO

ENTER [ Y/N ]: Y

ARCHIVE NORMAL IMAGE ON TAPE ?

Y = YES

N = NO

ENTER [ Y/N ]: Y

C1 : ^Z (tekan tombol EXIT CONTROL Z pada keyboard )

Tekan keypad MANUAL kemudian pilih teknik scanning sebagai berikut:

SCAN TIME – SEC : 8

mA : 25

kVp : 120

SOURCE FILTER : SHARP

SCAN DIAMETER – CM : 25

SCAN THICKNESS – MM : 2

Kemudian tekan keypad MANUAL SELECT dilanjutkan dengan keypad START SCAN, terlihat indicator LED pada START SCAN kedip-kedip (blinking) tunggu sampai berhenti sekitar 20 detik LED akan terus ON menandakan semua sistem sudah siap untuk melakukan proses scanning, lalu tekan keypad START SCAN, maka X-RAY TUBE di gantry akan membuka dan memancarkan sinarnya ke objek, sekitar 10 detik kemudian hasilnya berupa image muncul secara otomatis di monitor DISPLAY CONSOLE, dan menandakan proses scanning telah selesai.